Abstract

AbstractSpherically and irregularly shaped reversed phase packings were used to slurry pack capillary fused silica columns. The selection of the packing solvents was based on the colloidal properties of the stationary phase particles and investigated by sedimentation experiments. The chromatographic performance of the microcolumns was measured with conventional parameters from plate and rate theories, and the column resistance parameter and separation impedance. Also studied was the time of analysis. The performance of spherical and irregular packings was comparable with a light preference for spherically shaped materials when time of analysis is concerned. © 1995 John Wiley & Sons, Inc.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.