Abstract

abstract A series of shaking table and tilt tests were conducted with several commonly used strong-motion earthquake accelerographs. Table accelerations up to 0.5 g, both sine waves and recorded earthquake accelerograms, were used to evaluate response characteristics in the 1- to 50-Hz band under strong-motion conditions. All instruments performed within 10 per cent of manufacturers' specifications. No serious parasitic resonances were observed. Timing systems accuracies should be improved for intra-instrument detailed spectral comparisons. The general intuitive feel that mechanical, electronic and recording simplicity are a measure of expected performance for unattended instruments was reinforced by the tests.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.