Abstract

Aims and Objectives: To compare liver lesion conspicuity using torso phased-array (TPA) and body coils with two pulse sequences. Methods: Sixty patients with 125 focal hepatic lesions underwent T2-weighted fast spinecho (T2-FSE) and fast multiplanar inversion recovery (FMPIR) imaging with a standard body coil and with a TPA coil. The first 30 patients were scanned identically in both coils with four acquisitions; the second 30 were scanned with four acquisitions in the body coil and two in the TPA coil. Results: Liver-lesion contrast-to-noise (C/N) was significantly higher for the TPA coil both with four acquisitions ( P < 0.001) and with two acquisitions ( P < 0.002) using FMPIR, compared to with four acquisitions in the body coil. Liver-lesion C/N for T2-FSE was equivalent in both coils. Liver-lesion C/N was significantly higher ( P < 0.01) for FMPIR than T2-FSE both in the body coil and in the TPA coil. Conclusion: Liver-lesion C/N was significantly higher using the TPA coil rather than the body coil. Imaging time can be reduced by decreasing the number of acquisitions with the TPA coil.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.