Abstract

To compare effective dose differences when acquiring (1) dose reduction mode and (2) manual mode in a MORITA R100 CBCT. 24 exposure protocols with different technique factors were performed in both the dose reduction mode and the manual mode in a Veraviewepocs 3D R100 cone beam CT device. 48 TLD were placed in a RANDO head phantom at 24 different sites. Effective doses were calculated according to the formalism published in the 103rd ICRP report. Effective doses for the dose reduction mode protocols ranged from 14 to 156 µSv [mean = 57 µSv, standard deviation (SD) = 37 µSv], whereas effective doses for the manual mode protocols ranged from 22 to 267 µSv (mean = 94 µSv, SD = 65 µSv). Thus, across all protocols, the dose reduction mode leads to a drop of the effective dose by 38 % (SD = 6 %). The estimated effective doses are significantly lower if dose reduction protocols are acquired.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.