Abstract

The comparison between standard blasted and hydroxyapatite (HA)-coated cementless femoral stems (CLS stem) was made in the same patients who have bilateral late-stage osteonecrosis. Clinical and radiographic evaluations were performed in 20 patients (among 29 patients) with the minimum follow-up of 10 years (range, 123-168 months). Endosteal bone formation was found in all patients of both stems, and no radiolucent line was observed. The mean of initial stem subsidence was 1.27 ± 0.95 mm with the standard stem and 0.59 ± 0.56 mm with the HA-coated stem. Other results, such as hip score, endosteal bone formation, cortical hypertrophy, osteolysis, and heterotopic ossification showed no statistical differences. These have suggested that the fixation obtained by the cementless stem with a corundum-blasted surface is stable enough to get excellent clinical and radiographic results without the HA coating.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.