Abstract
The results for the comparison of Hp(0.07,0°) in 90Sr/90Y, carried out by the SSDL-ININ (pilot) and CPHR, are presented. Four calibration curves (CCs) are constructed with 32 TLD-200 dosimeters irradiated on a PMMA phantom. The CCs, TLD's Response RTLD (nC) vs Hp(0.07,0°), are in the range0.2-100 mSv. The SSDL-ININrate Hp(0.07, 0°) value is (43.85±0.57) µSv s-1 obtained with a primary extrapolation chamber serial 040. These same dosimeters are irradiated at the CPHR with a rate of Hp(0.07, 0°)=(14.94±0.36) µSv s-1, also calibrated with another primary standard extrapolation chamber serial 105. After irradiation at CPHR, the dosimeters are returnedto the SSDL-ININ for reading, and then irradiated again to construct the fourth CC (CC4). The comparison is based on a modification of the standard ANSI/HPS N13.11-2009, where the reference values are [Hp]ININ,ref. For the values of [Hp]CPHR the statistics bias B=0.002; standard deviation, S=0.099; and tolerance level L=0.099 are determined. The homoscedasticity test is performed for the variances associated to the mean performance quotientB.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.