Abstract

The reticulocyte hemoglobin content and hypochromic erythrocyte percentage offer advantages in evaluation of iron deficiency, especially in inflammatory conditions. The aim of this study was to evaluate the correlation of reticulocyte hemoglobin content (CHr, Ret-He) and hypochromic erythrocyte percentage (%HYPO, Hypo-He) between two automated hematologic analyzers. The CHr and %HYPO values were determined using the Advia 2021i (Siemens), while the Ret-He and Hypo-He levels were assessed using the XN-3000 (Sysmex). Data from a total of 971 cases and 834 patients were collected. For reticulocyte hemoglobin content, there was a good linear correlation between CHr and Ret-He (r = 0.857, p < 0.001). For percentage of hypochromic erythrocytes, there was a better correlation between the two measures when using a second-degree polynomial equation (Hypo-He* = 0.4818 - 0.0218 x %HYPO + 0.0069 x %HYPO2) (r = 0.786, p < 0.001).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.