Abstract

Polymeric coatings can provide temporary stability to bioresorbable metallic stents at the initial stage of deployment by alleviating rapid degradation and providing better interaction with surrounding vasculature. To understand this interfacing biocompatibility, this study explored the endothelial-cytocompatibility of polymer-coated magnesium (Mg) alloys under static and dynamic conditions compared to that of non-coated Mg alloy surfaces. Poly (carbonate urethane) urea (PCUU) and poly (lactic-co-glycolic acid) (PLGA) were coated on Mg alloys (WE43, AZ31, ZWEKL, ZWEKC) and 316L stainless steel (316L SS, control sample), which were embedded into a microfluidic device to simulate a vascular environment with dynamic flow. The results from attachment and viability tests showed that more cells were attached on the polymer-coated Mg alloys than on non-coated Mg alloys in both static and dynamic conditions. In particular, the attachment and viability on PCUU-coated surfaces were significantly higher than that of PLGA-coated surfaces of WE43 and ZWEKC in both static and dynamic conditions, and of AZ31 in dynamic conditions (P<0.05). The elementary distribution map showed that there were relatively higher Carbon weight percentages and lower Mg weight percentages on PCUU-coated alloys than PLGA-coated alloys. Various levels of pittings were observed underneath the polymer coatings, and the pittings were more severe on the surface of Mg alloys that corroded rapidly. Polymer coatings are recommended to be applied on Mg alloys with relatively low corrosion rates, or after pre-stabilizing the substrate. PCUU-coating has more selective potential to enhance the biocompatibility and mitigate the endothelium damage of Mg alloy stenting.

Highlights

  • Magnesium (Mg)-based alloys are bioresorbable scaffold materials that demonstrate enhanced properties of biodegradability and biocompatibility, and are being investigated for medical applications, such as atherosclerosis treatment [1, 2]

  • Endothelial cell attachment on samples was tested to evaluate the compatibility of Poly (carbonate urethane) urea (PCUU)- and PLGA-coated alloys in both static and dynamic conditions

  • PCUU coating on the surfaces of 316 L SS, WE43, AZ31, and ZWEKC improved the endothelial cell live/dead ratio in both static and dynamic conditions compared with non-coated alloys (P

Read more

Summary

Introduction

Magnesium (Mg)-based alloys are bioresorbable scaffold materials that demonstrate enhanced properties of biodegradability and biocompatibility, and are being investigated for medical applications, such as atherosclerosis treatment [1, 2]. Magnesium-based alloys have been studied in both in vitro and animal models [3,4,5], and Mg-based metal scaffolds have been tested in clinical trials for cardiovascular atherosclerosis [6]. There are few studies on magnesium alloys for other atherosclerotic diseases, such as intracranial atherosclerotic disease (ICAD), which has been claimed to be the most common cause of stroke worldwide [7]. Magnesium-based alloys are potentially biocompatible and provide high tensile strength (~542 MPa [10]) with lightweight properties, presenting new opportunities for cerebral stent application

Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.