Abstract

The objective of this study was to compare serial readings from an in–pharmacy automated blood pressure (BP) kiosk to mean daytime ambulatory BP. A total of 100 community–dwelling adults with hypertension underwent (1) three baseline automated office readings; (2) three in–pharmacy readings on each of four visits (12 total) using the PharmaSmart PS–2000 kiosk; and (3) 24–hour ambulatory BP monitoring between in–pharmacy visits two and three. Paired t–tests, Bland–Altman plots, and Pearson correlation coefficients were used for analysis. Mean BPs were 137.8 ± 13.7/81.9 ± 12.2 mm Hg for in–pharmacy and 135.5 ± 11.7/79.7 ± 10.0 mm Hg for daytime ambulatory (difference of 2.3 ± 9.5/2.2 ± 6.9 mm Hg [P ≤ .05]). Bland–Altman plots depicted a high degree of BP variability but did not show clinically important systematic BP differences. With ambulatory BP as the reference standard, in–pharmacy device results were similar to automated office results. The PharmaSmart PS–2000 closely approximated mean daytime ambulatory BP, supporting the use of serial readings from this device in the assessment of BP.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.