Abstract
We evaluated the performances of angiographic units equipped with a flat-panel (FP) detector and image intensifier (II) charge-coupled device (CCD) in the Interventional Cardiology (IC) Department. Entrance dose rate and dose per image, along with the dose at the II level were measured using 2 mm copper sheets to simulate a patient. Image quality (IQ) was evaluated using a phantom. Doses increased with fluoroscopy level changing from low to high. FP presented higher doses than CCD. Periodic measurements showed differences of up to 35%. Low mode IQ did not significantly differ from normal and high mode for both systems. Low fluoroscopy mode was decided to be used routinely. Both X-ray systems performed within international recommendations for conventional systems with the exception of higher cine radiation doses and II dose rates, stressing the fact that more studies are required to investigate whether dose levels should be adjusted.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.