Abstract
The radiation characteristics of a short source to surface distance (SSD) contact therapy tube in clinical use at the Centre Antoine-Lacassagne, Nice and a long SSD unit at the Clatterbridge Centre for Oncology were compared. The output from the tube at Nice had a dose rate of approximately double that of the tube at Clatterbridge, whereas the tube at Clatterbridge had a slightly higher value of the half value layer. Depth dose measurements were made with GafChromic MD55 film and SSD corrected depth dose curves showed good agreement between centres. Profiles at 2mm depth also showed comparable levels of flatness and uniformity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.