Abstract

The minimum in-stent lumen diameter is a predictor of restenosis. Stent dimensions provided by manufacturers are derived from in vitro tests. The aim of this study was to compare actual stent dimensions obtained by angiography and intracoronary ultrasound with dimensions that would be expected theoretically for a given inflation pressure in a cohort of 100 non-complex lesions suitable for direct stenting. Significant differences were found between the theoretical diameters and those observed by angiography and ultrasound. The actual-to-theoretical diameter ratio was 0.83 (0.09) when measured using angiography and 0.78 (0.10), using intravascular ultrasound. In lesions without severe calcification, stent dimensions were significantly smaller than indicated by the manufacturer. Nominal figures should not be used as reference values for stent implantation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.