Abstract
In order to enhance the resolution in a liquid scintillation with pulse shape discrimination for the measurement of actinide at low level, a study of two scintillating cocktails (Alphaex™ and Ultima Gold LLT™) and two alpha liquid scintillation spectrometers, (PERALS™ and TRI-CARB™) was undertaken. The criteria which can be applied to improve the resolution are the use of no-water miscible cocktail, the optimisation of energy transfer through the knowledge of the spectroscopic data, the employ of P.M. with higher quantum efficiency, scintillators with low Stoke's shift and high fluorescence quantum yield and the use of vial with low optical path.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.