Abstract

The use of Ir, W and Zr-coated graphite tubes, as permanent chemical modifiers, was studied for the determination of chromium in human scalp hair by ETAAS using the slurry sampling technique. The use of Mg(NO3)2 and Pd, as aqueous chemical modifiers, was also investigated and compared to the use of the permanent chemical modifiers. Scalp hair samples were pulverized using a Zr vibrational ball mill, (mean particle diameter of 0.8 μm) and suspended in ultrapure water. The lowest limit of detection, 44.5 μg kg–1,was achieved for the use of W-coated graphite tubes. The repeatability of the overall procedures (slurry preparation and ETAAS determination) were 15.7, 14.5 and 16.7% for W- and Zr-coated graphite tubes and Mg(NO3)2, respectively. The methods were applied to several reference materials, CRM 397 (human hair), DOLT-1 (dogfish liver) and DORM-1 (dogfish muscle), and the results obtained were in agreement to the certified values. Finally, the methods were applied to several human scalp hair samples from healthy adults.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.