Abstract

Monitoring of diabetes by measuring capillary blood glucose using a glucometer. To compare the three most used glucose meters in diabetic patients in a maternity and neonatal center in terms of repeatability, accuracy, and stability. A comparative study on 100 diabetic patients admitted to the various departments of a maternity and neonatal center. For each patient, a capillary measurement was made using each of the three glucometers to be tested (Accucheck®, On Call® and Bionime®) as well as a blood glucose on venous blood, performed in the laboratory using the Siemens X brand plus® PLC (reference method). The same sample was used to carry out all measurements. The Accucheck® brand reader and the On Call® brand thus, show a better correlation for open and closed bottles of test strips'storage, compared with the Bionime® brand reader. We found that the Accucheck® and On Call® readers are comparable in terms of reliability of results and are better than the Bionime® brand reader. We have found that, first using closed vials give more accurate results for glycemia readers and also, On Call® and Accucheck® brand readers are more reliable and accurate than the Bionime® brand reader.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.