Abstract

Currently, the most difficult-to-treat tibial plateau fractures are complex, comminuted fractures, which result from high-energy trauma. The anatomical reconstruction of the articular surface and the preservation of the soft tissue envelope are required for agood functional result. The aim of this study is to perform abiomechanical evaluation of three constructs used in the treatment of bicondylar fractures: the L‑shaped plate, the LCP locking plate, and the hybrid external fixator. We used 30experimental models divided into three study groups. There were ten models for each type of implant. The best biomechanical behavior was found with the LCP locking plate, which showed minimal deformation upon cyclic loading, followed by the hybrid external fixator. The LCP plate is the optimal implant for the stabilization of Schatzker type V tibial plateau fractures, with low rates of degradation for the implant and the bone-implant assembly.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.