Abstract

This study proposes a dark-field confocal microscopy (DFCM) based on fiber-mode excitation-assisted orbital angular momentum annular beam generation with a two-mode fiber to avoid diffraction distortion. The application of optical fibers compresses the DFCM volume, providing new ideas for system miniaturization. System adjustment difficulty is reduced and feasibility is verified by imaging 2D and 3D samples. High imaging contrast is achieved by fully blocking the reflected light and annular illumination with a steady central dark-spot within a propagation distance of 3 m. The application of our scheme can be further extended to detect subsurface defects in optical components and high-contrast biological imaging.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.