Abstract

A new method is presented that enables image acquisition to be segmented into two readouts. This is achieved using a new pulse sequence that creates two components of magnetization with different spatial profiles. Each component of the magnetization is measured in one of the readouts. This produces two images with complimentary "sensitivity profiles" and near identical contrast. The images can be acquired with a reduced data matrix that corresponds to shorter periods of data acquisition. The reduced matrix images are then combined to produce a full matrix image using reconstruction methods previously applied to images from multiple RF coils in the sensitivity encoding (SENSE) technique. The most promising application for this technique is in improving the performance of echo planar imaging (EPI) at high field. In this application, common SENSE obtains two segments of data in a single excitation of the magnetization (i.e., two readouts are performed per shot). The combination of these segments in image space avoids the difficulties normally associated with segmented EPI methods, namely, increased ghosting from discontinuities in the k-space data. The main advantages are a reduction in distortion and blurring. Common SENSE is compatible with parallel imaging and partial Fourier methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.