Abstract

Maintaining or even improving image quality while lowering patient dose is always the desire in clinical CT imaging. Iterative reconstruction (IR) algorithms have been designed to help reduce dose and/or provide better image quality. In this work, the channelized scanning linear observer (CSLO) is applied to study the combination of detection and estimation task performance using CT image data. The purpose of this work is to design a task-­‐based approach to quantitatively evaluate image-­‐quality for different reconstruction algorithms. Low-­‐contrast objects embedded in head-­‐size and body-­‐size phantoms are imaged multiple times and reconstructed by FBP and an IR algorithm for this study. Independent signal present and absent ROIs cropped from images are channelized by Difference of Gauss channels for CSLO training and testing. Estimation receiver operating characteristic (EROC) curves and the area under EROC curve (EAUC) are calculated by CSLO as the figure of merit. The One-­‐ Shot method is used to compute the variance of the EAUC values. Results suggest that the IR algorithm studied in this work could efficiently reduce the dose approximately 54% to achieve an image quality comparable to conventional FBP reconstruction for the combined detection and estimation tasks.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.