Abstract

We report a case of collapse of the Advanta V12 Large Diameter (LD) covered stent following previous successful implantation for native aortic coarctation. This unreported complication was successfully managed with subsequent stent redilation and implantation of two covered Cheatham-Platinum stents within the collapsed Advanta V12 stent. This case highlights the importance of correct stent apposition to the aortic wall and also raises some concerns regarding the radial strength of the Advanta V12 stent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.