Abstract

Coaxial time-resolved spectroscopy (TRS) based on a pump–probe technique using a color-selective double pulse (CSDP) is proposed. The CSDP, generated using an optical pulse shaper (OPS), was composed of different spectral components. Coaxially propagating CSDPs were used for pump and weak probe pulses. As a proof of concept, we evaluated the transient absorption of a ZnTe crystal from the difference between the temporal waveforms of the output chirped probe pulse with and without pump pulse using the OPS. Upon changing the pulse width of the probe pulse, the measured temporal width of the reaction was 0.2 ps, which agreed with that measured via conventional TRS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.