Abstract

A synthesized Schiff base ligand was studied to characterize its ability as a cation carrier in a poly(vinyl chloride) (PVC) membrane coated wire electrode. The constructed coated wire electrode showed excellent response characteristics to lead(II) ions. The electrode exhibits Nernstian slope of 29.1 mV/decade at 20 °C over a wide concentration range of Pb2+ ions from 1.5 × 10−6 to 1.3 × 10−1 M with a low detection limit of 9.0 × 10−7 M. The proposed potentiometric sensor has the following advantages: a very fast response, high reproducibility, long lifetime, and, most importantly, good selectivities relative to a wide variety of other cations. The selectivity behavior of the proposed Pb2+-selective electrode revealed a considerable improvement compared with the previously reported PVC membrane electrodes for lead(II) ions. The potentiometric response of the electrode is independent of the pH of the test solution in the pH range 3.5–7.5. The performance of the proposed electrode was investigated in partially non-aqueous media. The electrode was successfully applied to determine Pb2+ ions in mineral water, blood serum, and alloy samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.