Abstract

Objective: Examine the accuracy of automated audiometry in a clinically heterogeneous population of adults using the KUDUwave automated audiometer. Design: Prospective accuracy study. Manual audiometry was performed in a sound-treated room and automated audiometry was not conducted in a sound-treated environment. Study sample: 42 consecutively recruited participants from a tertiary otolaryngology department in Western Australia. Results: Absolute mean differences ranged between 5.12–9.68 dB (air-conduction) and 8.26–15 dB (bone-conduction). A total of 86.5% of manual and automated 4FAs were within 10 dB (i.e. ±5 dB); 94.8% were within 15 dB. However, there were significant (p < 0.05) differences between automated and manual audiometry at 250, 500, 1000, and 2000 Hz (air-conduction) and 500 and 1000 Hz (bone-conduction). The effect of age (≥55 years) on accuracy (p = 0.014) was not significant on linear regression (p > 0.05; R2 = 0.11). The presence of a hearing loss (better ear ≥26 dB) did not significantly affect accuracy (p = 0.604; air-conduction), (p = 0.218; bone-conduction). Conclusions: This study provides clinical validation of automated audiometry using the KUDUwave in a clinically heterogeneous population, without the use of a sound-treated environment. Whilst threshold variations were statistically significant, future research is needed to ascertain the clinical significance of such variation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.