Abstract

BackgroundIt is not rare for a small-bore feeding tube to be inserted incorrectly into the respiratory system in critically ill patients. Thus, monitoring is necessary to prevent respiratory malplacement of the tube. We investigated the utility of capnographic monitoring to prevent respiratory complications due to feeding tube mispositioning in critically ill patients.MethodsThis study was a pre and post-interventional study, including 445 feeding tube placements events studied retrospectively in the medical and surgical intensive care units of the Samsung Medical Center. We compared outcomes between time periods before and after capnographic monitoring and documented any respiratory complications.ResultsFeeding tubes were inserted in 275 cases without capnographic monitoring. Capnographic monitoring was performed in 170 cases. Sixteen patients (4%) had respiratory complications of all tube placements. Feeding tube was inserted into the trachea in 11 (2%) patients and for a pneumothorax in five (1%) patients. Fourteen cases of respiratory complications were detected in the control group (14/275, 5%, 10 tracheal insertions and four pneumothoraxes). Two respiratory complications were detected in the capnographic monitoring group (2/170, 1%, one tracheal insertion and one pneumothorax). Respiratory complications were detected less frequently in the capnographic monitoring group than that in the control group (P = 0.035).ConclusionsCapnographic monitoring is simple, easy to learn, and may be useful to prevent respiratory complications when placing a feeding tube in a critically ill patient.

Highlights

  • It is not rare for a small-bore feeding tube to be inserted incorrectly into the respiratory system in critically ill patients

  • Glasgow Coma Scale (GCS) was higher in the capnographic monitoring group than that in the control group (P = 0.005)

  • Two cases of respiratory complications occurred in the capnographic monitoring group (2/170, 1%, one tracheal insertion and one pneumothorax)

Read more

Summary

Introduction

It is not rare for a small-bore feeding tube to be inserted incorrectly into the respiratory system in critically ill patients. We investigated the utility of capnographic monitoring to prevent respiratory complications due to feeding tube mispositioning in critically ill patients. Small-bore feeding tubes have been used to provide proper nutrition and deliver medications to critically ill patients. Inserting a feeding tube is a routine procedure in many intensive care units (ICUs). Complications associated with inserting a feeding tube have been reported. Use of an assistant guidewire may contribute to respiratory complications when inserting a feeding tube [9,10,11].

Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.