Abstract

A new plate dialyzer, Gambro PRO-3, with a polycarbonate-polyether membrane (Gambrane) was tested. The membrane thickness is 16 micron and the dialyzer surface area 0.77 m2. 11 regular hemodialysis patients who were treated 3 times per week were studied and 32 dialyses were performed. Clearances of urea, creatinine, uric acid and inorganic phosphate at different blood flows were measured early and late in dialysis. Pulse rate and blood pressure were observed. Ultrafiltration rate and the residual blood volume were also measured. The performance of the dialyzer with regard to clearances was similar to that of the standard GLP 11.5 micron but inorganic phosphate clearance was slightly higher in PRO-3. The ultrafiltration of PRO-3 is about the same as for the GLP 11.5 micron dialyzer. No membrane leaks were observed, neither any adverse reactions in the patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.