Abstract

Low energy internal cardioversion is a safe and highly effective method for atrial fibrillation termination. We will describe 6 patients in whom the conventional 2-electrode systems with the defibrillation leads positioned in the right atrium and in the coronary sinus or left pulmonary artery failed to terminate the arrhythmia despite the use of maximal available energies. A 3-electrode configuration including right atrium, coronary sinus and left pulmonary artery was used in order to encompass as much atrial mass as possible between the cathode and the anode. The atrial fibrillation was successfully interrupted in 4 out of 6 patients. The creation of a 3-electrode configuration may be a further technical expedient in order to increase the success rate of internal cardioversion when usual manoeuvres like lead repositioning, reversion of polarity, or addition of antiarrhythmic drugs are ineffective.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.