Abstract

Laser scanning-based patient surface positioning and surveillance may complement image-guided radiotherapy (IGRT) as a nonradiation-based approach. We investigated the performance of an optical system compared to standard kilovoltage cone-beam computed tomography (CBCT) and its potential to reduce the number of daily CBCTs. We analyzed the patient positioning of 153 treatment fractions in 21 patients applied to three different treatment regions. Patients were first scanned with CBCT, shifted to the optimal isocenter position, and an optical scan was performed to verify the matching in relation to CBCT. For the head-and-neck region, the lateral/longitudinal/vertical/rotational/roll and pitch shift was 0.9 ± 1.8 mm/-2.7 ± 3.8 mm/-0.8 ± 3.6 mm/0.0 ± 1.1°/-0.5 ± 2.1°/0.2 ± 1.6°. For the thorax, the lateral/longitudinal/vertical/roll and pitch shift was -1.2 ± 3.6 mm/0.8 ± 5.1 mm/0.8 ± 4.3 mm/0.6 ± 1.4°/0.1 ± 0.9°/0.3 ± 1.0°. For the pelvis, the respective values were -2.5 ± 4.1 mm/4.6 ± 7.3 mm/-5.1 ± 7.4 mm/0.3 ± 1.1°/-0.5 ± 1.0°/0.3 ± 2.1°. In total, the recorded disagreement was -1.0 ± 3.6 mm/1.0 ± 6.3 mm/-1.8 ± 5.9 mm/0.3 ± 1.2°/-0.3 ± 1.5°/0.2 ± 1.7°. This analysis showed good agreement between the optical scanner approach and CBCT. The optical system holds potential to ensure precise patient positioning and reduced CBCT frequency in tumor locations with fixed relation to surface structures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.