The study was undertaken in order to determine clinical tactile evaluation of tetanic nerve stimulation (50 Hz, 5 sec) as a means of detecting fade and possible residual curarization. Forty-four patients were studied and 128 clinical evaluations by eight different observers were performed at various levels of neuromuscular blockade. Tetanic fade was detected reliably by clinical evaluation only when high degrees of fade (tetanic fade ratio of 0.3 or less) were detected with a force-displacement transducer. We conclude that the clinical tactile assessment of the response to tetanic stimulation is not accurate except at very high degrees of fade and that its usefulness in monitoring neuromuscular recovery is doubtful.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.