Preparation of ion-selective electrodes (ISEs) often requires long and complicated conditioning protocols limiting their application as tools for in-field measurements. Herein, we eliminated the need for electrode conditioning by loading the membrane cocktail directly with primary ion solution. This proof of concept experiment was performed with iodide, silver, and sodium selective electrodes. The proposed methodology significantly shortened the preparation time of ISEs, yielding functional electrodes with submicromolar detection limits. Moreover, it is anticipated that this approach may form the basis for the development of miniaturized all-solid-state ion-selective electrodes for in situ measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.