Abstract

We have described an electronically focused and scanned acoustic imaging device in earlier papers.1,2 This device was constructed by using a tapped acoustic surface wave delay line as a phase reference for an array of piezoelectric transducers. By inserting a frequency modulated chirp along the delay line, the device can be used either as a receiver or as a transmitter. In the transmit mode, it acts like a moving lens, emitting a focused beam which scans along a plane parallel to the array at a velocity comparable to the acoustic velocity. The same array can be used in the receive mode. By this means, a B-scan device was constructed which has good transverse and range definition..

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.