Abstract

The instantaneous frequency of a rapidly tuned continuous-wave (CW) laser is measured through linear optical sampling against dual-frequency combs. This dual-comb interferometer determines the instantaneous frequency of the CW laser during a quasi-sinewave frequency sweep of 3 THz amplitude with a 10 ms period. More complicated waveforms are also measured with instantaneous chirps exceeding 1500 THz/s (12 000 nm/s). The uncertainty is 1.5 MHz at 20 ns time resolution, averaging down to 5 kHz at 5 μs time resolution. The absolute frequency accuracy can be calibrated to within 2.5 kHz provided there is a brief period (<;1 ms) of low laser chirp (<;160 GHz/s) during the waveform measurement to allow for a dual-comb Vernier measurement of the absolute frequency, modulo 3 THz. This approach allows for the characterization of arbitrary CW waveforms with instantaneous frequencies that change rapidly and over wide optical bandwidth.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.