Abstract
An energy dispersive X-ray fluorescence (EDXRF) imaging system based on a Micropattern Gas Detector has already shown good results for different applications. An X-ray tube, a pinhole camera and a Micro-Hole and Strip Plate (MHSP) based detector are the main components of the experimental system. The detector uses an MHSP in a Xe atmosphere at 1 bar, and acting as a photon counting device, i.e., it is capable to record each single event retaining the energy and the interaction position (2D-sensitive detector) information of the incident photon, demonstrating to be a promising device for EDXRF imaging applications. This work presents studies of energy resolution, energy linearity and spatial resolution/elemental mapping as a function of image magnification of the system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.