An investigation of a high-pressure high-temperature (HPHT) synthetic diamond detector was performed with the aim to evaluate the potentiality of the detector for use in IMRT beams. Dosimetric parameters such as dynamics, stability of the response, linearity with dose and dose-rate dependence were studied and the HPHT sample behaviour was compared with that of a PTW natural diamond used as a reference system. In addition, a test in IMRT field using the step-and-shoot technique was also carried out on HPHT device. The main result of this study was the fast response shown by the HPHT sample, comparable to that of the natural diamond. Nevertheless, strong dose-rate dependence and the presence of the overshoot still limit the use of this system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.