Abstract

We characterized and calibrated a triple Gas Electron Multiplier (triple-GEM) detector, and studied its possible applications in medical dosimetry. The response to various sources of radiation was analyzed, and its gain, energy resolution, and time resolution were calculated. Then, radiation doses from an Iron-55 source and a medical portable X-ray machine were measured, obtaining the calibration factor for the GEM detector by comparison with reference values.We found an energy resolution of 19.5%, a time resolution of 40 ns, and a maximum gain of 5×106. The detector’s response to dose measurements was linear, with a calibration coefficient of 1.13×104 for different sources of radiation. It was concluded that GEM detectors can be reliably used as dosimeters in nuclear medicine and radiology.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.