Abstract

The calibration factor variation for a PTW natural diamond detector and a Scanditronix p-type stereotactic silicon diode (designed for use in photon beams) was studied in the 10–59 MeV range. Irradiations were performed in a water phantom with the 60 MeV ocular therapy beam at the CCO (UK). The diamond detector showed a sensitivity increase with energy, underestimating the dose by about 18% at the Bragg peak, by 7% at the centre and by 17% at the distal end of the SOBP region. The silicon diode did not show any significant sensitivity change with energy. However, a decrease in response of 24% was observed for an accumulated dose of 300 Gy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.