Abstract

We have developed a new detector using thin lanthanum bromide crystal (32 × 3 mm) for use in X-ray astronomy. The instrument was launched in high altitude balloon flight on two different occasions, December 21, 2007, which reached a ceiling altitude of 4.3 mbs and April 25, 2008 reaching a ceiling altitude 2.8 mbs. The observed background counting rate at the ceiling altitude of 4 mbs was ∼4 × 10 −3 ct cm −2 s −1 keV −1 sr −1. This paper describes the details of the experiment, the detector characteristics, and the background behaviour at the ceiling altitude.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.