Abstract

AbstractCharacterisation results from a single‐crystal CVD diamond detector are reported, which is under development for heavy‐ion particle identification and minimum‐ionizing particles timing in nuclear physics experiments. The charge collection efficiency is about 100%, never obtained from polycrystalline CVD‐diamond detectors. An energy resolution of 20 keV (ΔE/E ≈ 0.004) is achieved using a mixed nuclide α‐source, which is comparable to the energy resolution of silicon pin diode detectors. Using low impedance broadband electronics and a ToF technique, where holes or electrons drift separately inside the diamond bulk, the saturation velocity, mobility and lifetime of both charge carriers are estimated. In addition the thermoluminescent properties of the new material are discussed in comparison to polycrystalline diamond. A much lower density of trapping centres has been found for single‐crystal CVD diamond. (© 2005 WILEY‐VCH Verlag GmbH & Co. KGaA, Weinheim)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.