Abstract

Human hip joints with diseased cartilage and bone are commonly replaced using total hip replacements (THRs), which often have modular head and neck components that connect via matching taper, much like nested cones. Fretting corrosion occurs when subtle relative motion (i.e., micromotion) causes abrasion at the head–neck taper connection (i.e., fretting), thus allowing surrounding biofluids to degrade the material (i.e., corrosion). Corrosion then releases toxic metallic debris that causes local tissue lysis and painful inflammation, ultimately leading to revision surgery. However, concern with long-term systemic effects remains as the toxic debris diffuses throughout the body. Modular THRs are susceptible to fretting corrosion; however, they are important to allow intraoperative flexibility to reconstruct the joint with optimized biomechanics. Therefore, this chapter describes a procedure for investigating fretting corrosion of THRs, as well as how to analyze, present, and interpret results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.