Abstract

Four types of hydrosol filters, two reusable (diatomaceous cylinder and fritted-glass funnel) and two disposable (asbestos pad and membrane filter) were challenged with a heavy bacterial suspension to assess their ability to produce sterile filtrates. Two of the four diatomaceous earth filters, the four fritted-glass funnels, and all of the asbestos pads tested generally gave sterile filtrates. However, only one type of filter, one of the membranes in its manufacturer's own holder, consistently gave sterile filtrates. The two other types of membranes usually gave sterile filtrates if tested in one manufacturer's holder, but all types invariably gave contaminated filtrates when tested in another manufacturer's holder. Contaminated filtrates were generally attributed to a poor reusable filter or to a faulty holder used with a disposable filter. If a high degree of certainty is required for sterile heat-labile filtrate, it is suggested that the liquid be passed through two or more filters in a previously tested and proven system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.