Abstract

The optimal mode of femoral fixation in total knee arthroplasty (TKA) remains controversial, especially for the young patient. In a prospective randomised study we compared the magnitude and pattern of the fixation of cemented versus uncemented femoral components during 2 years in patients younger than 60 years. Forty-one knees in 41 patients were randomised to receive a NexGen (Zimmer, Warsaw, USA) cruciate-retaining TKA with either a cemented or an uncemented non HA-coated femoral component. The patients were examined by radiostereometric analysis (RSA), as well as clinical and radiological evaluation. The magnitude and pattern of migration as measured by RSA did not differ significantly between the cemented and uncemented fixation during the 2-year follow-up, nor were there any differences between the groups in clinical parameters. These findings suggest that an uncemented and non HA-coated femoral component may behave equally as well as a cemented one in the long-term.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.