Abstract
Implantable loop recorders (ILRs) allow continuous cardiac monitoring for 3-6 years and are a valuable tool for the investigation of syncopal episodes, palpitations, and atrial fibrillations as well as risk stratification after myocardial infarction. Although magnetic resonance imaging (MRI) in patients with ILRs has been shown to be safe, the impact of ILRs on cardiac MRI image quality has not been investigated yet. Thus, we tested the diagnostic value of cardiac MRI in patients with various types of ILRs. Two patients with an ILR and a clinical indication to assess myocardial burden of scarring and fibrosis or stress-induced myocardial ischemia underwent cardiac MRI. Device interrogation was performed prior to, immediately after, and 3 months after cardiac MRI. The post-MRI follow-ups revealed no change in programmed ILR parameters, sensing fidelity, and battery parameters. However, ILRs caused significant, uninterpretable hyperintensity artifacts in cardiac MRI. Further clinical studies are warranted to investigate whether modified MRI techniques are helpful to eliminate imaging artifacts.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.