Abstract

False-positive cardiac troponin I results as a result of carryover have previously been reported on the Beckman Coulter AccuTnI assay. We sought to determine if the carryover problem had been resolved with the new AccuTnI + 3 assay. Carryover experiments were performed in parallel on the Beckman Coulter Access2 analyser using the legacy AccuTnI and new AccuTnI + 3 assays. The same negative patient pool sample was analysed before and after a single analysis of an extremely elevated patient sample. Analysis of a single extremely high sample caused elevations above the 99th percentile cut-off, and thus false-positive cardiac troponin I results on both assays. Both assays demonstrated carryover and subsequent further elevations in negative pool results the following day. Our study replicates our previously published findings of carryover and reagent pack contamination on the AccuTnI assay. Despite improvements on the new AccuTnI + 3 assay, carryover and reagent pack contamination are still present.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.