Abstract

A technique is described for recording time concentration curves utilizing the thermal dilution method using normal room-temperature isotonic saline as the dilution indicator. This technique was compared with the dye dilution method by simultaneously recording the left heart output with each of the methods. The coefficient of correlation between these simultaneous measurements was 0.962. Paired analysis showed the thermal measurements to be 4% greater than dye measurements. However, their reproducibilities and sensitivities were comparable. Simultaneous thermal measurements for both sides of the heart resulted in a slightly higher reading (2%) from the left side. In spite of this small difference the reproducibilities with either side were comparable. The coefficient of correlation between the two was 0.96. A formula is advanced for calculating cardiac output from thermal dilution curves. The thermal dilution technique described provides a satisfactory method for estimating either right or left heart outputs. Submitted on August 12, 1960

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.