Abstract

The test of the inner tube integrity is an important checklist prior to the safe use of Bain's breathing system. This is because the major concern with the use of Bain's circuit is the potential malfunctioning of the circuit due to avulsion of the inner fresh gas delivery tube at the machine end which will turn the outer tube into dead space, a hazard recognized by Hannallah. Pethick test although widely used may not be foolproof to detect leaks in the inner tube. Ghani suggested the use of a plunger to perform the inner tube occlusion. Partially or completely occluded outer tube may produce a false positive result. Using very high flows or prolonged occlusion may cause damage to anesthetic machine due to high pressure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.