Abstract

A capillary viscometer designed for evaluating low-viscosity (0.2–10 cp) solutions at elevated temperatures [20–150 °C (68–302 °F)] has been built and extensively tested. This viscometer is automated, so that minimum operator effort is needed to obtain the viscosity as a function of temperature for a given solution. This viscometer is normally operated at a constant back pressure of 90 psi to avoid bubble formation. While the viscosities are routinely measured at a shear rate of about 38 s−1, viscosity measurement at other shear rates (10–250 s−1) can also be achieved by changing the set point on the flow controller.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.