Abstract

Capillary GC of metal chelates of diethyl dithiocarbamate (DDTC) was examined on a methylsilicone DB-1 column, (25 meter, 0.2 mm. i.d) with a film thickness of 0.25 μm. Elution was carried out at the initial column temperature of 180°C and programmed at 5°C min−1 to 260°C. Detection was by FID or ECD. Symmetrical peaks with base line separation were obtained with the metal chelates of copper(II), nickel(II), cobalt(III), manganese(II) and chromium(III). The ECD gave better sensitivity than the FID with a linear calibration range of 5–50 μg mL−1 and detection limits 2.0–6.0 μg mL−1, corresponding to 111–333 pg of metal ion reaching the detector. The method was applied to the determination of metal ions in water and pharmaceutical preparations with a coefficient of variation (CV) within 4.0%. When compared with a standard flame AAS method the results revealed no significant difference.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.