Abstract

The capillary electrophoresis method with direct UV detection is proposed for the determination of nitrite and nitrate in high-salt perchlorate solutions issued from uranium carbide dissolution. The isotachophoretic sample stacking was used to compensate for the perchlorate matrix interference. Simple electrolyte composed of 120 mM formiate buffer, pH 3.8 enabled the nitrate and nitrite determination in the presence of up to 1000-fold excess of perchlorate with 2 µM and 4 µM detection limits for nitrate and nitrite, respectively. The proposed method was applied to the determination of nitrate and nitrite in high-salt non-irradiated uranium carbide dissolution samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.