Abstract
In the present study, we have developed a capacitance electrochemical biosensor based on silicon nitride substrate (Si3N4/SiO2/Si[P]/Al) for Tumour Necrosis Factor Alpha (TNF-α) cytokines detection. Micro-contact printing, Fluorescence microscopy characterization and contact angle measurement (CAM) were carried out during the bio-functionalization of the biosensor surface. Mott-Schottky analyses were applied for TNF-α detection within the range of 1 pg/mL to 30 pg/mL in which the immunosensor has exhibited a good linearity, a sensitivity of 4 mV.pM−1 and 4.4 mV.pM−1 in PBS and artificial saliva (AS) respectively. While the LOD was found at 0.38 pg/mL and 1 pg/mL in PBS and AS respectively. The developed immunosensor has also demonstrated a high and good selectivity for TNF-α detection in human AS when compared to other interferences like Cortisol and Interleukin-10. The performances of the developed biosensor are very promising for biomedical application to predict the first sign of inflammation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.