Recently, investigators of otoacoustic emissions (OAEs) have shown interest in measuring OAEs to frequencies higher than 10 kHz. Most commercial instruments used to measure OAEs do not specify the microphone frequency response at higher frequencies, nor does their typically integrated design make it convenient to measure it. OAE probes manufactured by Etymotic Research have reasonably constant microphone sensitivity up to about 10 kHz and allow direct access to both the sound sources and microphone preamplifier output. A detailed procedure for calibrating the Etymotic Research OAE probe microphone to extend its usable frequency range to frequencies up to 20 kHz is described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.