Abstract

Abstract The efficiency of the Karlsruhe etched track detector was determined over a wide range of radon exposures. Sixty four detectors were placed in an airtight chamber. The radon concentration was kept at about 80 kBq.m-3 by circulating air over a solid radium source. At predetermined times, a group of four detectors was taken out of the chamber. The exposure period varied from a few hours up to several months in order to determine the efficiency curve for low exposure as well as for the region of saturation. The functional form y=a(1-e-bx) was fitted to the data. The low exposure range was fitted with only first six data points. The slope is 0.546 tracks.cm-2.kBq-1.m3.h-1, which is comparable to the earlier value of 0.55 tracks.cm-2.kBq-1.m3.h-1. The high exposure range was fitted with the whole data set. The two curves intersect at 2700 tracks.cm-2. At higher track densities the curve based on the whole data set has to be used to convert the number of tracks into an exposure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.